Typical adult ranges below are revision guides. Actual delivery depends on equipment, fit and breathing. Always check the manufacturer instructions and response. [1–5]
Nasal cannula
1–6 L/min · approximately 24–44% FiO₂Indications: Mild hypoxaemia; allows eating and speech.
Contraindications / limitations: Blocked/injured nose or insufficient oxygenation requires another device.
Side effects: Dryness, epistaxis, pressure injury at ears/nares.
Device care: Check prong placement and skin. FiO₂ changes with breathing pattern.
Simple face mask
5–10 L/min · approximately 35–60% FiO₂Indications: Moderate oxygen requirement in a spontaneously breathing patient.
Contraindications / limitations: Do not run below 5 L/min: CO₂ may accumulate. Avoid in vomiting or unprotected airway.
Side effects: Claustrophobia, skin pressure, CO₂ rebreathing at inadequate flow.
Device care: Keep exhalation ports open; arrange prescribed alternative support for meals.
Partial-rebreather mask
Usually 6–10 L/min · approximately 40–70% FiO₂; device dependentIndications: Selected reservoir-oxygen use according to local practice.
Contraindications / limitations: Not precise FiO₂; unsuitable for apnoea or inability to protect airway.
Side effects: Bag collapse, variable delivery, CO₂ rebreathing and pressure injury.
Device care: Inflate reservoir first; adjust flow to prevent collapse during inspiration. Follow device instructions.
Non-rebreather reservoir mask
10–15 L/min; commonly 15 L/min in emergencies · approximately 60–90% FiO₂Indications: Severe hypoxaemia while urgent assessment proceeds.
Contraindications / limitations: Does not ventilate an apnoeic patient. Poor seal or faulty valves reduces delivery.
Side effects: Supply failure, reservoir collapse, pressure injury and hyperoxia.
Device care: Pre-inflate reservoir; check valves and oxygen supply. Bag must not collapse on inspiration. Do not assume 100% FiO₂.
Venturi mask
Usually 24–60% FiO₂ · set flow printed on the selected adapterIndications: Controlled oxygen when a specified concentration is needed.
Contraindications / limitations: Do not obstruct entrainment ports or choose an arbitrary adapter/flow pairing.
Side effects: Discomfort, skin pressure; incorrect concentration if air ports are blocked.
Device care: Read adapter instructions; colour codes vary. Ensure total gas flow meets inspiratory demand.
Heated humidified HFNC
Adult systems commonly 20–60 L/min · FiO₂ 21–100%Indications: Selected acute hypoxaemic respiratory failure with close monitoring.
Contraindications / limitations: Cannot replace airway protection or delay intubation; major nasal/facial injury may prevent fitting.
Side effects: Discomfort, nasal injury, heat intolerance, possible air leak; delayed escalation.
Device care: Set flow and FiO₂ separately. Check correct-size prongs, heated humidifier, alarms and respiratory effort.
Tracheostomy collar / T-piece
FiO₂ set by blender/entrainment system; flow must meet demandIndications: Spontaneously breathing patient with an artificial airway.
Contraindications / limitations: No universal L/min-to-FiO₂ conversion. Blocked tube or apnoea needs airway/ventilation management.
Side effects: Dry secretions and obstruction, contamination, condensate aspiration.
Device care: Deliver to the stoma; provide appropriate humidification, assess tube patency and drain water away from patient.
Face tent
Variable FiO₂; commonly humidified flow ≥10 L/min, equipment dependentIndications: Selected mask intolerance or facial injury when a loose interface is suitable.
Contraindications / limitations: Not reliable for precise concentration or unstable severe hypoxaemia.
Side effects: Variable delivery, wet skin and contamination.
Device care: Check position and clinical response; follow manufacturer setup.
These provide respiratory support rather than a single fixed oxygen flow. FiO₂ and airway pressures are separately prescribed. NIV benefits selected COPD exacerbations with respiratory acidosis and cardiogenic pulmonary oedema. Arrest, inability to protect the airway, vomiting, major facial trauma or severe instability may require invasive airway management. Watch for mask injury, air leak, gastric distension, aspiration, hypotension and barotrauma. [6]
Adult settings must not be applied automatically to children. Nasal-prong settings depend on age, size and equipment. Hoods require adequate fresh-gas flow to clear CO₂ and measurement of concentration near the infant; one flow is not safe for every hood. WHO discourages routine head boxes/tents where efficient alternatives are available. Premature infants need neonatal targets to reduce hyperoxia-related injury. Nasal/nasopharyngeal catheters need trained insertion and local protocols. [7]